Unit 106, 7927 River Road, Delta, BC
Mon-Fri 09:00 AM - 05:00 PM
Unit 106, 7927 River Road, Delta, BC
Mon-Fri 09:00 AM - 05:00 PM
Shopping Cart
  • No products in the cart.
  • Listed Client – INTL-202119-C04

    CERTIFICATE OF COMPLIANCE

    Certificate Number: INTL-202119-C04 Project Number: INTL-202119-C04
    Date Issued: October 03, 2023 Client Number: INTL-202119
    Certification System: Type 3 Issued To: Beyes Dental Canada Inc.
    Unit 23, 595 Middlefield Road
    Toronto, ON, Canada M1V 3S2
    Certification Body: International Testing Laboratory Inc. Issued By: Certification Manager

    The products listed below are eligible to bear the INTL Certification Mark with adjacent indicator 'C' and ‘US’, and the subject area as shown below which represents that the product has been evaluated by an accredited CB to the applicable CSA and UL Standards respectively.

    Product: Panoramic, Cephalometric dental x-ray equipment
    Model #: Duray ART Plus, Duray ART Plus C
    Electrical Ratings: Duray Art Plus: 115Vac, 50-60Hz, Max 15A
    Duray Art Plus C: 115Vac, 50-60Hz, Max 15A
    Other Ratings: Ambient Temperature: 10 to 40 °C; Relative humidity: 10% to 75%; Pressure: 700 hPa to 1060 hPa
    Standards: CAN/CSA-C22.2 No. 60601-1:08
    ANSI/AAMI ES60601-1:2005
    CAN/CSA-C22.2 No.60601-1:08
    CSA C22.2 NO. 60601-2-63:15 (R2020)
    Conditions of Acceptability: - These units are intended to be used indoors environment within ambient temperature of Ambient Temperature: 10 to 40 °C; Relative humidity: 10% to 75%; Pressure: 700 hPa to 1060 hPa.
    - Not to be used in location classified as hazardous location.
    - Operating altitude: 2000m