CERTIFICATE OF COMPLIANCE
Certificate Number | INTL-202119-C02 | Project Number: | INTL-202119-C02 |
---|---|---|---|
Date Issued: | August 30, 2023 | Client Number: | INTL-202119 |
Certification System: | Type 3 | Issued to: | Beyes Dental Canada Inc. Unit 23, 595 Middlefield Road, Toronto, ON, Canada M1V 3S2 |
Certification Body: | International Testing Laboratory Inc. | Issued By: | Certification Manager |
The products listed below are eligible to bear the INTL Certification Mark with adjacent indicator 'C' and ‘US’, and the subject area as shown below which represents that the product has been evaluated by an accredited CB to the applicable CSA and UL Standards respectively.
Product: | Intra Oral Xray |
Model #: | Duray V2 X-ray Unit |
Electrical Ratings: | 110-127Vac; 50-60Hz; Max 8A, 0.95 KVA
220-240Vac; 50-60Hz; Max 5A, 1.1 KVA |
Other Ratings: | Ambient Temperature: 10 to 40 °C; Relative humidity: 10% to 90%; Pressure: 700 hPa to 1060 hPa
Transport and Storage Conditions: Ambient temperature: -20 to +50°C; Relative humidity: 10% to 90%; Pressure: 500 hPa to 1060 hPa |
Standards: | CAN/CSA C22.2 NO. 60601-1:14 (R2018)/UL 60601-1 (Ed.1).
CSA C22.2 NO. 60601-2-65:15 (R2020). |
Conditions of Acceptability: | - Intraoral Radiographic Equipment is intended to be used indoors environment within ambient temperature of -20 to +50°C; Relative humidity: 10% to 90%; Pressure: 500 hPa to 1060 hPa.
- Not to be used in location classified as hazardous location. - Operating altitude: 2000m |