CERTIFICATE OF COMPLIANCE
Certificate Number | INTL-202119-C03 | Project Number: | INTL-202119-C03 |
---|---|---|---|
Date Issued: | October 05, 2023 | Client Number: | INTL-202119 |
Certification System: | Type 3 | Issued to: | Beyes Dental Canada Inc. Unit 23, 595 Middlefield Road, Toronto, ON, Canada M1V 3S2 |
Certification Body: | International Testing Laboratory Inc. | Issued By: | Certification Manager |
The products listed below are eligible to bear the INTL Certification Mark with adjacent indicator 'C' and ‘US’, and the subject area as shown below which represents that the product has been evaluated by an accredited CB to the applicable CSA and UL Standards respectively.
Product: | Panoramic, Cephalometric and CBCT dental x-ray equipment |
Model #: | Duray 2D, Duray 2D CEPH, Duray 3D, Duray 3D CEPH |
Electrical Ratings: | Duray 2D: 115Vac, 50-60Hz, Max 16A
Duray 2D Ceph: 115Vac, 50-60Hz, Max 16A Duray 3D: 115Vac, 50-60Hz, Max 16A Duray 3D Ceph: 115Vac, 50-60Hz, Max 16A |
Other Ratings: | Ambient Temperature: 10 to 40 °C; Relative humidity: 10% to 75%; Pressure: 700 hPa to 1060 hPa |
Standards: | CAN/CSA C22.2 NO. 60601-1:14 (R2018)/UL 60601-1 (Ed.1).
CSA C22.2 NO. 60601-2-63:15 (R2020). |
Conditions of Acceptability: | - Panoramic and Cephalometric and CBCT dental x-ray equipment is intended to be used indoors environment within ambient temperature of 10 to 40 °C; Relative humidity: 10% to 75%; Pressure: 700 hPa to 1060 hPa.
- Not to be used in location classified as hazardous location. - Operating altitude: 2000m |